Biopure 200/300/600

Compliant, High Volume Water

The BIOPURE 200/300/600 reduces the contamination risk to patients by feeding high purity water at the point of use to large automated washer disinfectors and endoscope reprocessors.

Overview

It is specifically designed for sterile services/decontamination and endoscopy and is ideal for preventing the depositions commonly associated with using mains water to rinse endoscopes and automated endoscope reprocessors (AER).

The compact cabinet-style system can generate flows of 200, 300 or 600 litres of biopure water every hour and integrates easily with multiple washers from a wide variety of manufacturers, offering a unique opportunity for medical facilities to future-proof their washer systems.

 

Features and Benefits

The BIOPURE 200/300/600 cost-effectively provides biopure water at rates of 200 to 600 litres/hr.

 

Features

Benefits

Integral Class "A" Air Break-tank on feedwater supply

To comply with local by-laws and save space

High efficiency Reverse Osmosis membrane

For broad removal of organic/inorganics

Integral 350 litre biopure water reservoir with bacterial vent filter to maintain microbiological integrity and spay ball return to ensure full coverage of internal surfaces

To prevent biofilm build up

Variable speed recirculation pump

To provide instantaneous delivery flow whilst minimising temperature gain and noise levels

Ultra-violet steriliser

For continual bacteria destruction and maintenance of low microbiological levels

Ultrafiltration membrane

Providing final security against microbiological and particulate contamination

Integrated self-disinfection facility

To maintain the system in a hygienic state

Comprehensive monitoring and control system

To support validation and protect downstream processes by inhibiting bacterial growth opportunities

Hygienic sample point

To ensure reliable and consistent off line water quality monitoring

Small footprint (Depth and Width)

Maximises use of space

 

Applications

  • Sterile services
  • Decontamination
  • Endoscopy

 

Designed to meet:

  • HTM 01-06
  • NHS MES C32
  • EN 15883
  • MHRA/HIS Recommendations

Applications

Treated Water Specification

Model

Biopure 200

Biopure 300

Biopure 600

Make-up rate

200 l/h @ 15°C

300 l/h @ 15°C

600 l/h @ 15°C

Delivery flow rate

30 l/min @ 3.5 bar

30 l/min @ 3.5 bar

30 l/min @ 3.5 bar

Inorganic

<30 µS/cm*

<30 µS/cm*

<30 µS/cm*

Bacteria

<1cfu/10ml**

<1cfu/10ml**

<1cfu/10ml**

Endotoxin

<0.25Eu/ml

<0.25Eu/ml

<0.25Eu/ml

Particles

<0.05 micron

<0.05 micron

<0.05 micron

*Subject to a maximum feedwater conductivity of <700µS/cm. For use outside this range please contact our technical department

**System to be regularly sanitized and installed following our installation design guidelines. Sample to be taken from hygienic sample valve and sampled in accordance with our guidelines.

 

 

Feedwater Requirements

Contaminant

(Measure)

Range

Range

Range

Hardness

(Total Hardness)

(ppm as CaCO3)

<250

>250*

<10**

>10**

<10**

>10**

Total Chlorine

(CIppm)

<0.25

<0.25

<0.25

Silica

(Sio2ppm)

<5

<5

<5

Fouling Index (FI)

<10

<10

<10

Iron/Manganese

(Fe/Mn ppm)

<0.05

<0.05

<0.05

Temperature

Recommended

15-20°C

Recommended

15-20°C

Recommended

15-20°C

Flow rate

20 l/min

20 l/min

20 l/min

Drain requirements

 

Gravity fall with air gap, max during sanitization

45 l/min

45 l/min

45 l/min

Feedwater Pressure

10bar (150psi) max

2 bar (30psi) min

10bar (150psi) max

2 bar (30psi) min

10bar (150psi) max

2 bar (30psi) min

*Requires a softened water feed or an increase in the frequency of RO acid cleaning

**Requires a softened water feed as standard

 

 

Weights, Dimensions and installations

Height

1820mm (71.7")

1820mm (71.7")

1820mm (71.7")

Width

730mm (28.75")

730mm (28.75")

756mm (29.76")

Depth

890mm (35")

890mm (35")

1100mm (43.31")

Supply Weight

180kg (396lb)

180kg (396lb)

210kg (463lb)

Operational Weight

210kg (463lb)

210kg (463lb)

560kg (1,235lb)

Installation

Floor mounted

Floor mounted

Floor mounted

 

 

Electrical Requirements

Mains Input

230V ac, 50Hz

230V ac, 50Hz

230V ac, 50Hz

Systems control voltage

(not including pumps and UV)

24V dc

24V dc

24V dc

Power consumption (peak demand)

2200VA

2200VA

2200VA

Electrical protection rating

20amps

20amps

20amps

Noise level during recirculation

<60dBA

<60dBA

<60dBA

 

 

Self-Disinfection Program

Designed for ease of use by hospital staff, the Biopure incorporates a novel new disinfection system:

  • Accessed by secure E-Key control interface
  • Intelligent recognition of disinfectant concentration at start of cycle and during exposure phase and final rinse-out
  • Complementary Ultrafiltration, disinfectant and test strips to provide complete validation assurance
  • Ease of use
  • No removal of filters or disabling of UV lamp
  • Fully automated monitoring and control
  • User defined disinfection reminders